Intra-assay Precision (Precision within an assay): CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess.
Inter-assay Precision (Precision between assays): CV%<10%
Three samples of known concentration were tested in twenty assays to assess.
線性度:
To assess the linearity of the assay, samples were spiked with high concentrations of human CoQ10 in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay.
Sample
Serum(n=4)
1:20
Average %
96
Range %
91-101
1:40
Average %
99
Range %
93-104
1:80
Average %
92
Range %
87-96
1:160
Average %
90
Range %
85-94
回收率:
The recovery of human CoQ10 spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section.
Sample Type
Average % Recovery
Range
Serum (n=5)
84
80-88
EDTA plasma (n=4)
93
88-97
標(biāo)準(zhǔn)曲線:
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed.
Please provide the following information for Human Coenzyme Q10, CoQ10 ELISA Kit (cat:CSB-E14081h): 1.What is the standard? 2.What is the antibody in this kit? 3. Is the antibody against lipid Coq10 or to its binding protein? 4. will binding protein be labeled to Co1q10 or directly coq10?
A:
Thanks for your inquiry. The standard is purified CoQ10. This kit detects total CoQ10, and the antibody in the kit is mouse monoclonal antibody. The antibody is specific to CoQ10.Pls let me know if you have any further questions. Thank you.